Jay Vaishnav, PhD
Clinical Scientist, Manager, Medical Affairs Canon Medical Systems USA, Inc. |
Figure 1 Regions of interest extracted from images of the Catphan® 600 phantom low-contrast module, without (left) and with (right) PUREViSION technology (Aquilion ONE on the left; Aquilion ONE / GENESIS Edition on the right). Both images were obtained at similar CTDI values (left: 58 mGy, right: 59 mGy) and reconstructed with AIDR 3D. At similar CTDI, the PUREViSION image shows improved low-contrast detectability, with more of the simulated lesions visible on the PUREViSION images.
Figure 2 The introduction of AIDR 3D and FIRST iterative reconstruction has pushed imaging systems to ultra-low photon counts. This acquisition, with a nodule clearly visible in the axial view, was obtained at 2.5 CTDI, 98.9 DLP, or 1.38 mSv. PUREViSION technology reduces noise while maintaining high image quality.
Figure 3 This image shows a lung phantom imaged with conventional (left, Aquilion ONE / ViSION Edition) and PUREViSION (right, Aquilion ONE / GENESIS Edition) technology. Streak artifact reduction is visible in the inset. These are helical scans, obtained at the following parameters: Helical, 120 kV, AEC (SD7@5 mm, FC13, AIDR 3D Standard) , 0.5 s/r , 0.5 mm × 80, HP111, 350 mmΦ, FC03, AIDR 3D OFF, CTDIvol 6.7 mGy. Display parameters: WW=300, WL=30.
Figure 4 These images obtained without (left) and with (right) PUREViSION show similar image quality, at approximately half the CTDI; the left image was obtained at CTDI of 13.9 mGy; the right image at 7.2 mGy. The images were reconstructed at 5 mm slice thickness, with AIDR 3D Mild (FC 52) SD 10. Images courtesy of Fukuoka Wajiro Hospital, Japan.
Figure 6 CTA of aorta, without (Left) and with (Right) PUREViSION Optics.
Figure 7 A typical X-ray spectrum, showing the distribution of energies within a beam.
Figure 8 An X-ray tube generates photons at a range of energies (illustrated by the rainbow spectrum). Filtration removes the lower-energy components, leaving a more homogeneous beam.
Figure 9PUREViSION detector technology, from the Aquilion ONE / GENESIS Edition
Figure 10 Raw photon counts at detector vs. kVp, for PUREViSION vs. conventional filtration. The improved filtration not only results in a harder beam but in more photons reaching the detector, meaning more information that can contribute to the final image. At 80 kVp, the photon counts are approximately 40% greater.
Figure 11 Image noise for a conventional detector versus the PUREViSION detector, 40 cm diameter water phantom. Note that the PUREViSION detector demonstrates lower noise at all dose levels due to improved detector output. The noise reduction is more significant at lower doses due to a reduction in electronic noise. Lower noise translates to less need for signal and lower dose.
© Canon Systèmes médicaux Canada Limitée
© Canon Systèmes médicaux Canada Limitée